CelliGen 510

BioFlo/CelliGen 510 with screen_1
BioFlo/CelliGen 510 with screen_1
CelliGen 510
CelliGen 510

SIP benchtop reactors designed to meet the needs of R&D through production. More information

Product Information

The CelliGen 510 is a compact sterilizable-in-place cell culture system ideal for pilot to small scale production processes. The modular design features a wide variety of standard options allowing this system to be flexible enough to meet most process requirements.

Gas flow
› Various combinations of one, two, three or four TMFCs:
0.06  –   3 SLPM
0.3  –   15 SLPM
0.6  –   32 SLPM
1  –   64 SLPM
› Air Wash System with TMFC (0  – 15 or 0  –  32 SLPM)
› Overlay with TMFC (0 –  15 or 0  – 32 SLPM)
› Overlay (valve only)

Exhaust line
› Exhaust condenser
› Manual or automatic pressure control

Impellers
› Rushton-type – Used commonly for robust cell lines such as bacteria, yeasts, and algae, where maximum OTR is desired.
› Pitched-blade – Commonly used with mammalian, insect or other shear sensitive cell lines for batch, fed-batch or continuous cultures. Produces axial and radial mixing. Provided as standard with the CelliGen 510 bioreactors.
› Marine – Commonly used with mammalian, insect, or other shear sensitive cell lines for batch, fed-batch or continuous cultures. Produces axial mixing
› Spinfilter – For suspension or microcarrier cultures where a dip tube inside the filter withdraws cell media as harvest or waste
› Cell-Lift – Specially designed to provide uniform circulation for both suspension and microcarrier cultures. Can be used with optional decanting columns for perfusion cultures
› Packed-bed – For secreted products from anchorage-dependent and suspension cells, this impeller immobilizes cells in a bed of Fibra-Cel® Disks to provide extremely high cell densities

Sensors
› Single or redundant pH/DO
› Redox
› Foam/level

Addition and sampling
› Resterilizable sampling valve
› Resterilizable addition valves (4 max)
› Resterilizable addition/harvest valve with dip tube (2 max)
› 1.5 in sanitary fitting 7-port septum
› Addition vessels (glass/stainless steel)
› Decanters
› Sterile sampling assembly

Validation
› Basic Package
› Basic Plus Package
› Enhanced Package

Utility regulator & pre-filter kits
› Process air pre-filter/regulator kit
› Instrument air pre-filter/regulator kit
› Water pre-filter/regulator kit
› Process steam regulator kit
› Utility steam pre-filter/regulator kit
› Utility connection kit

Additional options
› Sprayballs for clean-in-place of vessel
› External variable-speed pumps
› External scales
› Spare parts kits
› Preventative maintenance kits
› Mobile table
› System passivation
› Glycol/chiller heat exchanger

Features

  • RPC (Reactor Process Control) software
  • 32 L stainless-steel electropolished vessels with 2:1 aspect ratios
  • Integrated load cell
  • 3-Fixed speed peristaltic pumps for the additions, harvest, and pH control
  • Benchtop or Mobile option
  • Choice of pitched-blade, marine, spin filter, cell-lift or packed-bed impellers
  • industrial touchscreen interface

Applications

  • Laboratory- and pilot-scale fermentation of aerobic and anaerobic bacteria, yeasts, and fungi
  • Laboratory- and pilot-scale cell culture of mammalian, insect, and human cell lines
  • Special applications such as stem cell culture, vaccine, monoclonal antibody or biofuel/biopolymer production
  • Suitable for batch, fed-batch, and continuous/perfusion processes
CelliGen 510
Video: CelliGen 510 – Cell Culture Impellers
CelliGen 510
Video: CelliGen 510 – BioCommand Setup